Related Products

Denntal High-Speed LED Handpiece L-A 110234

Standard Configuration

High-Performance LED High-Speed Air Turbine Handpiece

Key Features

Built-in Generator & LED Light: Provides natural and bright illumination for precision.

Enhanced Durability: Copper body with a special German dynamic balancing process for stability.

Optimized Performance: High-torque chuck mechanism for improved efficiency and easy bur replacement.

Superior Cooling & Low Noise: Three-point spray ensures full-range cooling, reducing heat and discomfort.

User-Friendly Design: Quick coupling system for effortless installation and operation.

Specifications

Working Air Pressure: 0.25 MPa – 0.30 MPa

Speed Range: 320,000 – 420,000 RPM

Bur Compatibility: φ1.593mm – φ1.598mm

Noise Level: ≤63 dB

Interface: ISO-E International Standard

Featured of Dental LED Highspeed HandPiece L-A 110234

Check Our All Products

Precision Engineered for Excellence

The L-A 110234 High-Speed Handpiece is designed for superior performance, ensuring smooth and stable operation. It delivers high-speed rotation for precision dental procedures, making it an essential tool for professionals seeking reliability and efficiency.

High-Speed Performance for Efficient Treatment

With an impressive speed range of 320,000 – 420,000 RPM, this high-speed handpiece ensures efficient and precise cutting, reducing chair time while optimizing clinical outcomes.

Built-in Generator & LED Illumination

The handpiece features an integrated generator that powers a high-quality LED light, providing bright and natural illumination. This enhances visibility and improves accuracy during procedures, ensuring optimal treatment precision.

Superior Cooling & Low Noise

Equipped with an advanced three-point spray system, the handpiece delivers efficient cooling, reducing heat and ensuring a more comfortable experience for both the patient and the operator. Additionally, its low-noise performance (≤63 dB) minimizes disturbances during procedures.

Quick Coupling System for Easy Installation

Designed for convenience, the handpiece incorporates a quick coupling system, enabling effortless installation and disassembly. This feature allows for easy maintenance and seamless instrument changes during dental treatments.

Universal 2-Hole & 4-Hole Compatibility

The handpiece supports both 2-hole and 4-hole connection types, ensuring broad compatibility with various dental unit systems, making it a versatile choice for different clinical setups.

Dental Handpiece Different Connecting Type:

Check Our All Products

Normal handpiece :how to match tightly handpieceair tube and chair connector

Check Our All Products

Dental High-Speed Handpiece Cartridge Disassembly and Installation Guide

Check Our All Products

Disassembly:

1:Align the wrench cap position with your left thumb on the cap.

2:Turn counterclockwise to open the cap.

3:Use your index finger to gently move out the cartridge.

Installation:

1:Ensure that the cap has a spring pad in place.

2:Position the head of the cartridge in the cap (as shown in the image), and with the hand holding the bur, ensure it fits smoothly.

3:Turn the cap clockwise a few times, then tighten using the wrench.

Installation Guide

This simple process ensures easy and secure cartridge replacement for optimal handpiece performance.

Changing Burs for High-Speed Handpiece

Check Our All Products

1: With your right hand, hold your thumbs up and press the gland to open the chuck.

2: Use your left hand to carefully remove the bur.

3: To load a new bur, follow the same steps as above, ensuring the bur is properly aligned with the shaft hole.

This ensures secure and accurate placement of the bur for efficient use during procedures.

Dental High-Speed Handpiece Maintenance

1: Apply special lubricating oil before and after each use.

2: After use, blow air through the head and bur transfer position to clean them. Tilt the handpiece at a 45-degree angle to avoid blowing debris inside the head.

3: Blow air through the two small tubes at the tail to remove residual moisture and debris.

4: Once a month, remove the cartridge, and clean any foreign bodies from the cavity in the head and cap.

Proper maintenance helps ensure the longevity and optimal performance of your handpiece.

Dental High-Speed Handpiece Exception Occurs And Solution

  • Problem:Noisy
  • Reason : 1: Bearing damage; 2: Bur curved; 3: cap without cushion
  • Solution: 1: Change bearing or bur; 2: Put new washer

  • Problem: Light slowly or does not ligh
  • Reason : 1: Poor water quality to block the water hole 2: Atomization gas too much to hold back the water
  • Solution: 1: Use a silver needle to drill the bearing hole (as the picture shows)
    2: The ato water or no wamization will be small much

  • Problem: Poor power
  • Reason : 1.Foreign body to block the head into the hole ; 2.Chair exhaust pipe plug exhaust is not suitable
  • Solution: 1.using the file into the air hole; 2.Vent the air to the vent holesuitable

  • Problem: Can not catch the bur
  • Reason : 1.Bur is not standard; 2.Spindle is damaged
  • Solution: 1.Change standard bur; 2.Change spindle

  • Problem: Leakage
  • Reason : 1.Connection is not tightening; 2.O-ring is damaged
  • Solution: 1.Connect with the Chair twist; 2.Change plug o-ring

  • Problem: Connections twist tightly
    Reason : The Chair joints damage
    Solution: Change chair Connector

What Sets Lavender Dental Apart

Lavender Dental is a trusted name in the dental industry, renowned for providing high-quality, reliable dental equipment to professionals worldwide. Our brand stands out for its commitment to excellence, innovative solutions, and dedication to customer satisfaction. Whether you're outfitting a new dental practice or upgrading your existing equipment, Lavender Dental is your go-to partner for superior products and unmatched expertise.